Guidance and manufacture’s declaration-electromagnetic immunity
for EQUIPMENT and SYSTEMS that are not LIFE-SUPPORTING
Guidance and manufacture’s declaration-electromagnetic immunity
The PO80 Pulse Oximeter is intended for use in the electromagnetic environment specified below. The customer or the user of PO80 Pulse
Oximeter should assure that it is used in such an environment.
Immunity test IEC60601 test level Compliance
level
Electromagnetic environment -guidance
Radiated RF
ICE 61000-4-3
3V/m
80MHz to 2.5GHz 3V/m
Portable and mobile RF communications equipment should be used
no closer to any part of the PO80 Pulse Oximeter, including cables,
than the recommended separation distance calculated from the
equation applicable to the frequency of the transmitter.
recommended separation distance
Where P is the maximum output power rating of the transmitter in
watts (W) according to the transmitter manufacturer and d is the
recommended separation distance in meters (m).
Field strengths from fixed RF transmitters, as determined by an elec-
tromagnetic site survey,
a
should be less than the compliance level in
each frequency range
b
Interference may occur in the vicinity of equipment marked with the
following symbol:
NOTE 1 At 80MHz and 800MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is aected by absorption and reflection from structures,
objects and people.
a
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio,
AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment
due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which The
PO80 Pulse Oximeter is used exceeds the applicable RF compliance level above, the PO80 Pulse Oximeter should be observed to verify
normal operation. If abnormal performance is observed, additional measures may be necessary, such as reorienting or relocating the PO80
Pulse Oximeter.
b
Over the frequency range 150 KHz to 80 MHz, field strengths should be less than 3V/m.
Recommended separation distances between portable and mobile
RF communications equipment and the EQUIPMENT or SYSTEM
for EQUIPMENT or SYSTEM that not LIFE-SUPPORTING
Recommended separation distances between
portable and mobile RF communications equipment and the PO80 Pulse Oximeter
The PO80 Pulse Oximeter is intended for use in the electromagnetic environment in which radiated RF disturbances are controlled. The cus-
tomer or the user of the PO80 Pulse Oximeter can help prevent electromagnetic interference by maintaining a minimum distance between
portable and mobile RF communications equipment (transmitters) and the PO80 Pulse Oximeter as recommended below, according to the
maximum output power of the communications equipment.
Rated maximum output
power of transmitter
(W)
Separation distance according to frequency of transmitter
(m)
80MHz to 800MHz 800MHz to 2.5GHz
0.01 0.117 0.233
0.1 0.369 0.738
1 1.167 2.333
10 3.689 7.379
100 11.67 23.33
For transmitters rated at a maximum output power not listed above, the recommended separation distanced in meters (m) can be estimated
using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W)
according to the transmitter manufacturer.
NOTE 1 At 80MHz and 800MHz, the separation distance for the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is aected by absorption and reflection from structu-
res, objects and people.
Mat. Nr. 750.803 - 0114 Irrtum und Änderungen vorbehalten
4


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Beurer PO 80 at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Beurer PO 80 in the language / languages: English, German as an attachment in your email.

The manual is 4,02 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

Others manual(s) of Beurer PO 80

Beurer PO 80 User Manual - Dutch - 32 pages

Beurer PO 80 User Manual - French - 32 pages

Beurer PO 80 User Manual - Italian - 32 pages

Beurer PO 80 User Manual - Spanish - 32 pages


The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info