Device care
Clean the device only with a soft, dry cloth.
Cleaning the cuff
Take out the bladder. Fold and place the cuff cover inside a washing
bag. Wash the cuff cover with warm water and a mild detergent in
a washing machine. Air dry the cuff. DO NOT iron the cuff cover.
Accuracy test
The law recommends to verify the static calibration of a blood
pressure monitor every two years or after mechanical shock.
How to check device calibration:
−Switchthisdeviceintothecalibrationmode(CAmode)byre-
pressing the Start button while the device memory is displayed.
−Connectareferencepressuresourcebyuseofthedevice
calibration connector (available as a special accessory)
−Comparethedisplayedpressurewiththeexternalreference
pressure
−Forfurtherinformationpleasecontactyourlocaldealer.
Disposal
Battery and electronic devices must be disposed of in
accordance with the locally applicable regulations, not with
domestic waste.
8.Guarantee
This device is covered by a 2 year guarantee
from the date of
purchase.
•Battery,andwearingpartsarenotincluded.
•Thecuffisincludedforthefunctionalguaranteeonly(tightness)
for 1 year.
•Openingoralteringthedeviceinvalidatestheguarantee.
Theguaranteedoesnotcoverdamagecausedbyimproper
handling, discharged batteries, accidents or non-compliance with
the operating instructions.
Please contact your dealer (see foreword).
9. Technical Specifications
Operating condition:
10-40°C/50-104°F
15-90%relativemaximumhumidity
Storage condition: -20-+55°C/-4-+131°F
15-90%relativemaximumhumidity
Weight: 157.7g(includingbattery)
Dimensions: 167x60x41mm
Measuring procedure:
oscillometric, corresponding to Korotkoff
method Phase I systolic, Phase V
diastolic
Measurement range:
30-280mmHg–bloodpressure
40-200beatsperminute–pulse
Cuff pressure display
range: 0-299mmHg
Resolution:
1 mmHg
Static accuracy: pressure within ± 3 mmHg
Pulse accuracy: ±5%ofthereadoutvalue
Voltage source:
 specifybatterytype(3.7V,350mAh)
rechargeable via standard MicroUSB
Battery life: Approximately200measurementsper
Reference to
standards:
recharge cycle
ANSI/AAMI/ISO81060-2;
EN1060-1/-3/-4;
IEC60601-1;IEC60601-1-2(EMC);
IEC 60601-1-11
Expected service life: 5yearsor10,000measurements
(battery and cuff are not included)
IP Class: IP20
ThestipulationsofEU-Directive93/42/EECfor
Medical Devices Class IIa have been fulfilled
7


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Microlife VSA at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Microlife VSA in the language / languages: English as an attachment in your email.

The manual is 0.58 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info