www.medicalsuction.co.uk
5
MAINTENANCE
The NEW ASKIR 30 12V suction equipment does not need maintenance or lubrication.
It is, however, necessary to inspect the unit before each use. With regard to training, given the information contained in the user manual
and since it is easy to understand the said device, it doesn’t appear to be necessary.
Unpack the instrument and always check integrity of plastic parts and feeding cable, they might have been damaged during previous
use. Connect the cable to electrical network and turn the switch on.
Close the aspirator outlet with your finger and with suction regulator at maximum check that the vacuum indicators reaches at least -
75kPa (-0.75 bar). Rotate the knob from right to left. The vacuum indicator should go down -25kPa (-0.25 bar).
Check that no loud noises are present. The device is protected by a safety fuse (F 10A L 250V) situated in the cigarette lighter adapter.
For fuses replacing, always the type and the range.
Before changing the fuse, disconnect the plug from the power supply socket.
Fault type Cause Solution
1.The suction unit doesn’t work
Cable is damaged
External power source failure
Replace the cable
Check the external power source
2.No aspiration
Jar Cap badly screwed down Unscrewed the cap, then rescrew it correctly
3. No aspiration Lid seal not in its seat Unscrew the cap and insert the seal properly in its seat
4.The Vacuum power on the
patient side is either very low
or absent
a) Vacuum regulator set to minimum
b) Protection filter blocked or
damaged
c) Connection tubes blocked, kinked
or disconnected
d) Shut-off valve blocked or
damaged
e) Pump motor damaged
a) Turn the vacuum regulator clockwise and check the
value of the vacuum on the gauge
b) Replace the filter
c) Replace or reconnect the tubes, check the jar
connections
d) Empty the jar, or disconnect the tube from the jar and
unblock the shut-off valve. The unit twill only work in the
upright position
e) Refer to authorised service personnel
5. The float doesn’t close If the cap has been washed, ensure
that the float is not partially detached
Insert the float into it’s place
6. The float doesn’t close The float it’s covered by dirty
material
Unscrewed the cap, leave the and put in on autoclave
7. Low suction Foam inside the jar Fill the jar to 1/3 full of ordinary water
Faults 1 - 2 - 3 - 4 - 5 - 6 - 7 None of the remedies has
achieved the desired results
Contact the seller or CA-MI After-sales Assistance
Service
If the overfill security system it’s activated, don’t proceede with the liquid aspiration.
If the overfill security system doesn’t work there are two cases:
case If the overfill security system doesn’t work the aspiration will be stopped by the bacteriological filter who avoid the liquid
penetration inside the device.
case If both the security system doesn’t work, there is the possibility that liquid comes inside the device, in this case return the
device to CA-MI technical service.
CA-MI Srl will provide upon request electric diagrams, components list, descriptions, setting instructions and any other information that
can help the technical assistance staff for product repair
.
BEFORE EVERY CHECKING OPERATION, IN CASE OF ANOMALIES OR BAD FUNCTIONING, PLEASE CONTACT CA-MI
TECHNICAL SERVICE. THE MANUFACTURER DOES NOT GIVE GUARANTEE IF INSTRUMENT, AFTER THE TECHNICAL
SERVICE CHECKING, APPEARS TO BE TAMPERED.
INSTRUCTIONS
The device must be checked before each use in order to detect malfunctions and / or damage caused by transport and / or
storage.
The working position must be such as to allow one to reach the control panel and to have a good view of the empty indicator, the
jar and the antibacterial filter.
It is recommended not to keep the device in your hands and / or to avoid prolonged contact with the body of apparatus.
WARNING: For proper use, place the aspirator on a flat, stable surface in order to have the full volume of use of the jar and better
efficiency of the overflow device.
Place the unit on a flat, horizontal surface
Connect one end of the short silicon tube, with antibacterial filter, to the suction connector on the lid of the jar.
The other tube already connected to the filter has to be connected to the “VACUUM” jar outlet, where has been fixed the red float
(security float). When the 90% of the volume of the jar is reached there is the activation of the security float (the float close the
aspiration connector on the jar) to avoid liquid penetration inside the device.
6


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Cami New Askir 30 at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Cami New Askir 30 in the language / languages: English as an attachment in your email.

The manual is 0,31 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info