www.medicalsuction.co.uk
4
ACCESSORIES SUPPLIED
DESCRIPTION
COMPLETE ASPIRATION JAR 1000cc
CONICAL FITTING
TUBES SET 6 mm x 10 mm
ASPIRATION PROBE CH20
ANTIBACTERIAL FILTER
Antibacterial Filter: The filter is produced with (PTFE) hydrophobic material witch prevents fluids entering the pneumatic circuit.
The filter is for a single patient use which will protect patients and machines from cross contamination.
When the filter is wet, it’s not possible to use the unit therefore the filter should be changed immediately.
In case of possible contamination or discolouration, change the filter immediately.
Don’t use the suction unit without the protection filter fitted.
Suction catheter: Single-use device to be used on a single patient. Do not wash or re-sterilize after use. Reuse may cause cross-
infections.
Don’t use after lapse of the sell-by date
WARNING: Suction tubes for insertion in the human body purchased separately from the machine should comply with ISO
10993-1 standards on material biocompatibility.
Aspiration jar: The mechanical resistance of the component is guaranteed up to 30 cycles of cleaning and sterilization. Beyond this
limit, the physical-chemical characteristics of the plastic material may show signs of decay. Therefore, we recommend that you to
change it.
Silicone tubes: the number of cycles of sterilization and/or cleaning is strictly linked to the employment of the said tube. Therefore, after
each cleaning cycle, it is up to the final user to verify whether the tube is suitable for reuse. The component must be replaced if there
are visible signs of decay of the material constituting the said component.
Conical fitting: the number of cycles of sterilization and the number of cleaning cycles is strictly linked to the employment of the said
component. Therefore, after each cleaning cycle, it is up to the final user to verify whether the fitting is suitable for reuse.
The component must be replaced if there are visible signs of decay in the material constituting the said component.
Service life of the device: More than 850 hours of operation (or 3 years) in accordance with the standard conditions of testing and
operation. Shelf life: maximum 5 years from the date of manufacture.
CLEANING OF ACCESSORIES
To clean the plastic housing of the device wear disposable latex gloves and clean with denaturated alcohol or hypoclorite solutions.
Washing and / or cleaning the autoclavable jar as to be carried out as follows:
Wear protection gloves and apron (glasses and face mask if necessary) to avoid contact with contaminating substances;
Disconnect the jar from the device
Disconnect all tubes from the jar and the protection filter
Empty and dispose of the content and of the suction catheter according to the laws in force in your country;
Separate all parts of the cover (overflow valve, o-ring);
After disposing of disposable parts and disassembling the jar wash in running cold water and rinse thoroughly.
Then soak in warm water (temperature shall not exceed 60°C).
Wash thoroughly and if necessary use a non-abrasive brush to remove incrustations.
Rinse in running warm water and dry all parts with a soft cloth (non-abrasive).
The jar and the cover can be autoclaved by placing the parts into the autoclave and running one sterilization stem cycle at 121°C
(1 bar relative pressure) making sure that the jar is positioned upside down.
Mechanical resistance of the jar is guaranteed up to 30 cycles of sterilization and cleaning at the indicated conditions
(EN ISO 10079-1). Beyond this limit the physical-mechanical characteristics of the plastic may decrease and replacement of the part is
therefore recommended.
After sterilization and cooling at environment temperature of the parts make sure that these are not damaged.
Assemble the jar as follows:
Place the overflow valve into its seat in the cover (under VACUUM connector)
Insert floating valve keeping the o-ring towards the opening of the cage
Place the o-ring into its seat around the cover
After completing assembling operations always make sue that cover seals perfectly to avoid vacuum leakages or liquid exit
The aspiration tubes can be sterilized on autoclave using a sterilization cycle at 120°C.
The conical connector can be sterilized on autoclave using a sterilization cycle at 121°C.
The device is ready for a new employment now.
DO NOT WASH, STERILIZE OR PUT IN AUTOCLAVE THE ANTIBACTERIAL FILTER
5


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Cami New Askir 30 at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Cami New Askir 30 in the language / languages: English as an attachment in your email.

The manual is 0,31 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info