Guidance and manufacturer’s declaration – Electromagnetic Emissions
The surgical aspirator NEW ASKIR 30 12V is intended for use in the electromagnetic environment specified below.
Emissions Test Compliance Electromagnetic environment -
guidance
Irradiated / Conducted emissions
CISPR11
Group 1
The surgical aspirator NEW ASKIR 30 12V
only used RF energy only
for its internal functioning. Therefore, its
RF emissions are very low and
are not cause interference in proximity of
any Electronic appliances.
Irradiated / Conducted emissions
CISPR11
Class [B]
The surgical aspirator NEW ASKIR 30 12V
can be used in all
environments, including domestic and
those connected directly to the
public mains distribution that supplies
power to environments used for
domestic scopes
Harmonic emissions IEC/EN 61000-3-2 Class [A]
Voltage fluctuations / flicker emissions
IEC/EN 61000-3-3
Complies
Guidance and manufacturer’s declaration Electromagnetic Immunity
The surgical aspirator NEW ASPIRET is intended for use in the electromagnetic environment specified below.
The customers or the user of the surgical aspirator NEW ASPIRET should assure that it’s used in such an environment.
Immunity Test Compliance Electromagnetic environments -
guidance
Electrostatic discharge (ESD)
IEC/EN 61000-4-2
± 6kV on contact
± 8kV in air
Floors should be wood, concrete or ceramic
tile. If floors are covered with synthetic
material,
the relative humidity should be at least
30%.
Electrical fast transient / burst
IEC/EN 61000-4-4
± 2kV power supply
Mains power quality should be that of a
typical
commercial environment or hospital
Surge
IEC/EN 61000-4-5
± 1kV differential mode
Mains power quality should be that of a
typical
commercial environment or hospital
Loss of voltage, brief voltage interruptions
and variations
IEC/EN 61000-4-11
5%U
T
for 0.5 cycle
40%U
T
for 05 cycle
70%U
T
for 25 cycle
<5%U
T
for 5 sec
Mains power quality should be that of a
typical commercial environment or hospital
If
the user of the surgical aspirator NEW
ASPIRET.
request that the appliance operates
continuously, the use of a continuity unit is
recommended.
Magnetic field
IEC/EN 61000-4-8
3A/m
The power frequency magnetic field should
be
measured in the intended installation
location to
assure that it’s sufficiently low.
Conducted Immunity
IEC/EN 61000-4-6
3Vrms 150kHz to 80MHz
(for appliances that aren’t life - supporting)
-
Irradiated Conducted
IEC/EN 61000-4-3
3V/m 80MHz to 2.5 GHz
(for appliances that aren’t life - equipment)
-
Note U
T
is the value of the power supply voltage
4


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Cami New Askir 30 at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Cami New Askir 30 in the language / languages: English as an attachment in your email.

The manual is 0,31 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info