5. Probe LED Specifications
Wavelength Radiant Power
RED 660±2nm 1.8mW
IR 940±10nm 2.0mW
6. Power Requirements
Two AAA alkaline Batteries
Power consumption: Less than 30mA
Battery Life: Two AAA 1.5V, 800mAh alkaline batteries could be continuously operated as long as 28 hours.
7. Environment Requirements
Operation Temperature: 5
~40
Storage Temperature: -20
~+55
Ambient Humidity: 80% no condensation in operation; 93% no condensation in storage
Atmosphere pressure: 86kPa~106kPa
8. Equipment Response Time
As shown in the following figure.
Response time of slower average is 12.4s.
9. Classification
According to the type of protection against electric shock: INTERNALLY POWERED EQUIPMENT;
According to the degree of protection against electric shock: TYPE BF APPLIED PART (The application part is rubber inside of the
Pulse Oximeter);
According to the degree of protection against ingress of water: IPX1
According to the mode of operation: CONTINUOUS OPERATION
FCC Declaration
This device complies with Part 15 of the FCC Rules. Operation is subject to the following two conditions:
(1) This device may not cause harmful interference, and
(2) This device must accept any interference received, including interference that may cause undesired operation.
Please take attention that changes or modification not expressly approved by the party responsible for compliance could void the
user’s authority to operate the equipment.
Note: This product has been tested and found to comply with the limits for a Class B digital device, pursuant to Part 15 of the FCC
Rules. These limits are designed to provide reasonable protection against harmful interference in a residential installation. This
p
roduct generates, uses, and can radiate radio frequency energy and, if not installed and used in accordance with the instructions,
may cause harmful interference to radio communications. However, there is no guarantee that interference will not occur in a
particular installation. If this product does cause harmful interference to radio or television reception, which can be determined by
turning the equipment off and on, the user is encouraged to try to correct the interference by one or more of the following measures:
—Reorient or relocate the receiving antenna.
Increase the separation between the equipment and receiver.
—Connect the equipment into an outlet on a circuit different fro
m that to which the receiver is connected.
—Consult the dealer or an experienced radio/TV technician for help.
Declaration
Guidance and Manufacturer’s declaration electromagnetic emissions-For all EQUIPMENT and SYSTEMS
Guidance and Manufacturer’s declaration - electromagnetic emission
The SMG01 Pulse Oximeter is intended for use in the electromagnetic environment specified below. The customer or the user
of SMG01 Pulse Oximeter should assure that it is used in such an environment.
Emission test Compliance Electromagnetic Environment – guidance
RF emissions CISPR 11 Group 1 The SMG01 Pulse Oximeter uses RF energy only for its
internal function. Therefore, its RF emissions are very low
and are not likely to cause any interference in nearby
electronic equipment.
RF emissions CISPR 11 Class B
Harmonic emissions
IEC 61000-3-2
Not Applicable
Voltage fluctuations/ flicker emissions
IEC 61000-3-3
Not Applicable
The pulse Oximeter (SMG01) is suitable for use in all
establishments, including domestic establishments and
those directly connected to the public low-voltage power
supply network that supplies buildings used for domestic
purposes.
Guidance and Manufacturer’s declaration electromagnetic immunity-For all EQUIPMENT and SYSTEMS
Guidance and Manufacturer’s declaration - electromagnetic immunity
The SMG01 Pulse Oximeter is intended for use in the electromagnetic environment specified below. The customer or the user
of the SMG01 Pulse Oximeter should assure that it is used in such an environment.
Immunity test IEC 60601 test level Compliance
Level
Electromagnetic Environment – guidance
Electrostatic
Discharge (ESD)
IEC 61000-4-2
+/- 6kV contact
+/- 8kV air
+/- 6kV contact
+/- 8kV air
Floors should be wood, concrete or ceramic tile. If
floor are covered with synthetic material, the relative
humidity should be at least 30%.
Power frequency (50/60
Hz) magnetic field
IEC 61000-4-8
3A/m 3A/m Power frequency magnetic fields should be at levels
characteristics of a typical location in a typical
commercial or hospital environment.
Guidance and Manufacturer’s declaration electromagnetic immunity-
For all EQUIPMENT and SYSTEMS that are not LIFE-SUPPORTING
Guidance and Manufacturer’s declaration - electromagnetic immunity
The SMG01 Pulse Oximeter is intended for use in the electromagnetic environment specified below. The customer or the user
of the SMG01 Pulse Oximeter should assure that it is used in such an environment.
Immunity
test
IEC 60601
test level
Compliance
Level
Electromagnetic Environment – guidance
Radiated
RF
IEC
61000-4-3
3 V/m
80 MHz to
2.5 GHz
3 V/m
Portable and mobile RF communications equipment should be used no closer
to any part of the Pulse Oximeter (SMG01), including cables, than the
recommended separation distance calculated from the equation applicable to
the frequency of the transmitter.
Recommended separation distance
80 MHz to 800 MHz 800 MHz to 2.5 GHz
Where P is the maximum output power rating of the transmitter in watts (W)
according to the transmitter manufacturer and d is the recommended
separation distance in meters (m).
Field strengths from fixed RF transmitters, as determined by an
electromagnetic site survey
a
, should be less than the compliance level in
each frequency range.
b
Interference may occur in the vicinity of equipment marked with following
symbol:
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations, Electromagnetic propagation is affected by absorption and reflection
structures, objects and people.
a Field strengths from fixed transmitters, such as base station for radio (cellular/cordless) telephones and land mobile radios,
amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the
electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the
measured field strength in the location in which the Pulse Oximeter (SMG01) should be observed to verify normal operation. If
abnormal performance is observed, additional measurements may be necessary, such as reorienting of the relocating the Pulse
Oximeter (SMG01).
b Over the frequen
cy range 150 kHz to 80 MHz, fields strengths should be less than 3 V/m
Recommended separation distances between portable and mobile RF communications equipment and
the EQUIPMENT or SYSTEMS - For all EQUIPMENT and SYSTEMS that are not LIFE-SUPPORTING
Recommended separation distances between
portable and mobile RF communications equipment and Pulse Oximeter (SMG01)
The Pulse Oximeter (SMG01) is intended for use in electromagnetic environment in which radiated RF disturbances are
controlled. The customer or the user of the Pulse Oximeter (SMG01) can help prevent electromagnetic interference by
maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the Pulse
Oximeter (SMG01) as recommended below, according to the maximum output power of the communications equipment.
Separation distance according to frequency of transmitter (m) Rated maximum output
power of transmitter (W)
80 MHz to 800 MHz
800 MHz to 2.5 GHz
0.01 0.1167 0.2334
0.1 0.3689 0.7378
1 1.1667 2.3334
10 3.6893 7.3786
100 11.6667 23.3334
For transmitters rated at a maximum output power not listed above, the recommended separation distanced in meters (m) can
be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of
the transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and refle ction
from structures, objects and people.
Possible Problems and Solutions
Problems Possible reason Solution
SpO
2
or PR can not
be shown normally
1. Finger is not inserted correctly
2. Patient’s SpO
2
value is too low to be measured
1. Retry by inserting the finger
2. There is excessive illumination
3. Try some more times. If you can make sure no
problem exist in the product, please go to a
hospital timely for exact diagnosis.
SpO
2
or PR is shown
unstably
1. Finger might not be inserted deep enough.
2. Excessive patient movement
1. Retry by inserting the finger
2. Be calmness
The oximeter cannot
be powered on
1. No battery or low power of battery
2. Batteries might be installed incorrectly
3. The oximeter might be damaged
1. Please replace batteries
2. Please reinstall the batteries
3. Please contact with local customer service
centre
Indication lamps are
suddenly off
1. The product is automatically powered off when no
signal is detected longer than 8 seconds
2. The battery power is too low to work
1. Normal
2. Replace the batteries
“Error3” or
“Error4” is displayed
on screen
1. Err 3 means the red emission LED is damaged
2. Err 4 means the infra-red emission LED is damaged
1. Check the red emission LED
2. Check the infra-red emission LED
Error6
Err 6 means the screen is failure Please contact with local customer service centre
“Error7” is displayed
on screen
Err 7 means all the emission LED or reception diode is
damaged.
Please contact with local customer service centre
Symbol Definitions
Symbol Definition Symbol Definition
Type BF applied part.
Attention
Protected against dripping water.
SpO
2
Oxygen saturation
PR bpm
Pulse rate (BPM)
Low power indication
No SpO
2
Alarm
SN
Serial No.
Storage temperature and relative
humidity
Follow instruction for use
Manufacturer’s information
Date of Manufacture
Bluetooth indication
Indicate the signal is not stable
Notes:
1. The illustrations used in this manual may differ slightly from the appearance of the actual product.
2. The specifications are subject to change without prior notice.
Baros GmbH
Kleine Düwelstr. 21
30171 Hannover
0123
2


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for A-Rival oQsy at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of A-Rival oQsy in the language / languages: English as an attachment in your email.

The manual is 2,8 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

Others manual(s) of A-Rival oQsy

A-Rival oQsy User Manual - German - 2 pages


The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info