Accuracy is always the biggest concern of a thermometer user. Rossmax ThermoCal is a portable
and convenient device that allows the user to easily check the accuracy of a thermometer in just a
few seconds. The innovative design of the adjustable blackbody cavity applies to all three types of
Rossmax thermometer products - pen type, ear and forehead thermometer.
Parts name/location
Control panel Back view
Introduction
Instructions for installation
Cautions and maintenance
Specications
Operation procedures
1. Check the accuracy
Step 1: Check water level of the bath.
Step 2: Turn on the device.
Step 3: Select a testing temperature on the control panel (default at 37°C).
Step 4: Wait for the water to reach the desired temperature.
Step 5: Choose the blackbody cavity corresponding to the thermometer types.
Step 6: Set the thermometer into the calibration mode.
Step 7: Insert the probe/tip into the cavity.
Step 8: Start the measurement.
Step 9: Thermometer will beep when the measurement is completed.
Step 10: Compare the reading given by the thermometer to the calibration temperature.
The acceptable allowance is within ±0.2°C.
2. Drain the water bath
Step 1: Make sure the device is off.
Step 2: Remove the tube from the upper end.
Step 3: Empty the water bath by lowering the upper end of the tube.
Step 4: Plug the tube back in after the draining is completed.
1. Install the device on a solid surface.
2. Make sure the power is off before plugging in the device.
3. Keep the device away from the environment with direct sunlight, high temperature, high
humidity or severe change in temperature.
4. Make sure the AC power and its grounding plug are properly connected.
5. Do not use power extension cords.
6. Use pure water or distilled water only for the water bath to avoid contamination or water
heater breakdown.
7. Pay attention to the water level to avoid overspill.
8. The water bath is to remain 90% full at all times to reach water temperature stability.
1. Make sure the device, AC power and its grounding plug, is plugged in properly.
2. Change water periodically, once a month is suggested.
3. Use pure water or distilled water only for the water bath. Other water sources might increase
the chance of calcium carbonate formation on the outside of the water heater.
4. Make sure there is enough water in the water bath every time before using the device. Please
constantly pay attention to the water level and hygiene when/if the device runs for over 6 hours
or the water temperature exceeds 40°C.
5. Keep the device away from physical impact.
6. Unplug the device and empty the water bath if the device is going to be inactive for a long
time.
7. Unplug the device before moving it or draining the water bath.
8. Clean the water bath with a 3M scrub sponge if any signs of iron oxidization appear.
9. Pay attention to the water temperature to avoid scald.
Power source AC 100-240V 50/60 Hz
Power of water heater 255W
Testing temperature Ambient temperature + 5°C to maximum 45°C
Ambient temperature 16 ~ 30°C
Stability of water bath temperature ± 0.02°C
Diameter of
blackbody cavity
Pen type 15 mm
Ear 12.5 mm
Forehead 55 mm
Capacity 5.39 Liters
Weight 9.69 kgs
Dimension 180(W) x 350(H) x 320(D) mm
Control
panel
Ventilation
holes
Water
bath actual
temperature
Water bath set
temperature
Heating
indicator
Temperature up
Temperature down
Digit selecting key
Temperature setting key
Socket
and fuse
Power
switch
Cooling
fan
Overheating
protection
setting knob
Drain
tube
Removable
bath cover
Blackbody
cavity
Computer
2


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Rossmax ThermoCal - Professional Testing Device for Thermometer at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Rossmax ThermoCal - Professional Testing Device for Thermometer in the language / languages: English as an attachment in your email.

The manual is 0,95 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info