2. Heating indicator of the water bath does not light up after the calibration temperature is set
Possible causes Actions to take
Water temperature is higher than the set
temperature
Cool down the water in the bath
Defective heating indicator Contact Rossmax for service
Defective temperature controller Contact Rossmax for service
Set temperature is too low
Set temperature should be at least 5°C higher
than operating temperature
3. Heating indicator of water bath lights up, whereas the water temperature remains low
Possible causes Actions to take
Defective heater circuit Contact Rossmax for service
Defective temperature controller Contact Rossmax for service
4. Water temperature continues to rise and exceed the set temperature
Possible causes Actions to take
Defective temperature controller Contact Rossmax for service
5. Large uctuation in water temperature
Possible causes Actions to take
Water temperature is not stable yet Wait for temperature to stabilize
Set temperature is too close to ambient
temperature
Lower the ambient temperature
Defective temperature controller Contact Rossmax for service
Defective water heater Contact Rossmax for service
6. Power and power indicator are both on, whereas the water circulating pump does not start
Possible causes Actions to take
Foreign object clogs inside the pump Remove the foreign object
Loose screw used to secure the propeller Tighten the screw
Defective motor Contact Rossmax for service
7. Noise generated by the stirrer motor
Possible causes Actions to take
Bearing ball oxidation Replace bearing
Bearing aging Replace bearing
Trouble shooting
1. Power is switched on but the power indicator does not light up
Possible causes Actions to take
No power input Check the power source
Loose plug Insert the plug into the socket securely
Device is powered and running but the power
indicator does not light up
Replace the power switch
Power wire broke or loose connection Replace the power wire(s)
Tripped fuse Check the fuse and replace it if needed
WARNING: The symbol on this product means that it’s an
electronic product and following the European directive
2012/19/EU the electronic products have to be disposed
on your local recycling centre for safe treatment.
1


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Rossmax ThermoCal - Professional Testing Device for Thermometer at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Rossmax ThermoCal - Professional Testing Device for Thermometer in the language / languages: English as an attachment in your email.

The manual is 0,95 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info