549543
21
Zoom out
Zoom in
Previous page
1/28
Next page
21
Troubleshooting
Trouble Possible cause Corrective action
The nebulisation rate is
extremely low.
Batteries are low (the orange
battery low indicator flashes).
Replace the batteries with new
ones. Refer to page 9.
The mesh is stained or clogged. 1. Clean the mesh by boiling. Refer
to page 16.
2. Replace the mesh cap with a new
one. Refer to page 11.
The green power
indicator does not turn
on and the unit does not
nebulise.
The batteries are low. Replace the batteries with new
ones. Refer to page 9.
The capacity of rechargeable
batteries is low.
Charge the batteries with a
commercially available recharger.
The batteries are inserted in the
wrong direction.
Re-insert the batteries in the correct
direction. Refer to page 9.
The AC adapter is not correctly
connected to the main unit.
Remove the AC adapter from the
main unit and connect it correctly.
The medication container is not
installed correctly to the main unit.
Remove the medication container
from the main unit and install
correctly. Refer to page 12.
The green power
indicator turns on
but the unit does not
nebulise.
The vibrator is not immersed in the
medication.
Tilt the main unit in order to immerse
the vibrator in the medication.
The medication container is not
filled with medication.
Fill with medication. Refer to page
12.
The mesh is broken. Replace with a new one. Refer to
page 11.
21


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Omron NE-U22 - MicroAir at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Omron NE-U22 - MicroAir in the language / languages: English as an attachment in your email.

The manual is 3,34 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

Others manual(s) of Omron NE-U22 - MicroAir

Omron NE-U22 - MicroAir User Manual - German - 28 pages

Omron NE-U22 - MicroAir User Manual - Dutch - 28 pages


The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info