704396
1
Zoom out
Zoom in
Previous page
1/8
Next page
5104-001-08
© 2019 Nonin Medical, Inc.
MPS, Medical Product Service GmbH
Borngasse 20
D-35619 Braunfels, Germany
Nonin and Onyx are registered trademarks of Nonin Medical, Inc.
U.S. Patents 5,490,523; 5,792,052
EC
REP
Nonin Medical, Inc.
13700 1st Avenue North
Plymouth, Minnesota 55441, USA
+1 (763) 553-9968 (outside US and Canada)
+31 (0)13 - 79 99 040 (Europe)
(800) 356-8874 (US and Canada)
Fax: +1 (763) 553-7807
+31 (0)13 - 79 99 042 (Europe)
E-mail: info@nonin.com
infointl@nonin.com (Europe)
nonin.com
Instructions for Use—English
Onyx
®
II Model 9550 Finger Pulse Oximeter
Indications for Use
The Nonin
®
Onyx II Model 9550 Finger Pulse Oximeter is a small, lightweight, portable device indicated for use in
measuring and displaying functional oxygen saturation of arterial hemoglobin (%SpO
2
) and pulse rate of patients
who are well or poorly perfused. It is intended for spot-checking of adult and pediatric patients on fingers (other than
the thumb) between 0.3 – 1.0 inch (0.8 – 2.5 cm) thick. The device’s intended use environments include hospitals,
clinics, long-term care facilities, skilled nursing facilities, emergency medical services, and home healthcare
services.
Warnings
Do not use the device in an MR environment, in an explosive atmosphere, or on infant or neonatal patients.
This device is not defibrillation proof per IEC 60601-1.
Inspect the sensor application site at least every 4 hours to ensure correct sensor alignment and skin integrity. Patient
sensitivity to the sensor may vary due to medical status or skin condition.
Avoid excessive pressure to the sensor application site as this may cause damage to the skin beneath the sensor.
This device is intended only as an adjunct in patient assessment. It must be used in conjunction with other methods of
assessing clinical signs and symptoms.
The device must be able to measure the pulse properly to obtain an accurate SpO
2
measurement. Verify that nothing is
hindering the pulse measurement before relying on the SpO
2
measurement.
Operation of this device below the minimum amplitude of 0.3% modulation may cause inaccurate results.
General operation of the device may be affected by the use of an electrosurgical unit (ESU).
The use of accessories other than those specified in these instructions may result in increased electromagnetic emission
and/or decreased immunity of this device.
This device should not be used adjacent to or stacked with other equipment. If adjacent or stacked use is necessary, the
device should be observed carefully to verify normal operation.
Keep the oximeter away from young children. Small items such as the battery door, battery, and lanyard are choking hazards.
Certain activities may pose a risk of injury, including strangulation, if lanyard should become wrapped around your neck.
Before changing batteries, make sure the device is off and is not applied to a digit.
Portable RF communications equipment such as cell phones or radios (including peripherals such as antenna cables and
external antennas) should be used no closer than 30 cm (12 inches) to any part of the ME system, including cables
specified by the manufacturer. Otherwise, degradation of the performance of this equipment could result.
Cautions
This device has no audible alarms and is intended only for spot-checking.
This device is designed to determine the percentage of arterial oxygen saturation of functional hemoglobin. Factors that
may degrade pulse oximeter performance or affect the accuracy of the measurement include the following:
The device may not work when circulation is reduced. Warm or rub the finger, or re-position the device.
This device’s display will go blank after 30 seconds of no readings or poor readings.
In some circumstances, the device may interpret motion as good pulse quality. Minimize patient motion as much as possible.
Clean the device before applying it to a patient.
Do not sterilize, autoclave, or immerse this device in liquid. Do not pour or spray any liquids onto the device.
Do not use caustic or abrasive cleaning agents, or any cleaning agent containing ammonium chloride.
This device is a precision electronic instrument and must be repaired by Nonin Technical Service. Field repair of the device
is not possible. Do not attempt to open the case or repair the electronics. Opening the case may damage the device and
void the warranty.
A flexible circuit connects the two halves. Do not twist or pull the flexible circuit or overextend the device’s spring. Do not
hang the lanyard from the device’s flexible circuit.
A functional tester cannot be used to assess the accuracy of a pulse oximeter monitor or sensor.
II
do not apply the pulse oximeter on the
same arm as a blood pressure cuff,
arterial catheter or infusion line(s) (IVs)
excessive light, such as sunlight or
direct home lighting
excessive motion
moisture in the device
improperly applied device
finger is outside recommended size
range
poor pulse quality
venous pulsations
anemia or low hemoglobin
concentrations
cardiogreen and other intravascular
dyes
carboxyhemoglobin
methemoglobin
dysfunctional hemoglobin
artificial nails or fingernail polish
residue (e.g., dried blood, dirt, grease,
oil) in the light path
1


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Nonin Onyx II - 9550 at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Nonin Onyx II - 9550 in the language / languages: English as an attachment in your email.

The manual is 2,37 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info