Insulation of measurement leads
Insulation of measurement leads may deteriorate over time. This can cause
measurement and/or safety related problems. If insulation material leaks current
measured results will be wrong. Also in such case there is a risk of electric shock when
the user touches deteriorated part of insulation.
Eurocheck is not needed to perform this test.
Step Action
1 q Switch on the tested instrument and select insulation measurement with
1000 V test voltage (highest available).
2 q Connect universal test cable to the instrument. Make sure that all
measurement leads are in direct connection to each other over the
whole length.
q Leave all leads open.
3 q Start insulation measurement. The instrument must signal over-range
measurement result.
4 q If the instrument measures lower-than-maximum insulation, stop
using this measurement cable immediately.
Note:
q It is possible to verify other types of measurement cables as well. Just connect
them to the instrument and perform insulation measurement as described above.
4.14 PE test terminal operation verification
Warning:
q This test uses voltages that are hazardous live. Only adequately trained
personnel may perform this test. Ensure that disconnected lead is not exposed.
Verification can be performed with universal test cable only.
Step Action
1 q Connect tested instrument into Eurocheck RCD/Fault loop output using
universal test cable (use test tips).
q Connect only PE and N leads into L and N sockets respectively.
2 q Switch on the tested instrument and select fault loop impedance, line
impedance or RCD function.
3 q Connect Eurocheck into mains wall socket and enable RCD/Fault loop
output.
q If Eurocheck signals incorrect L and N polarity, reverse connection.
4 q Touch PE terminal on the tested terminal with your finger – the
instrument must signal dangerous line situation.
q Write down test result for documentation purposes.
5 q If test fails the tested instrument is damaged. Send it immediately
to your distributor or manufacturer for repair.
22


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Metrel Eurocheck CS 2099 at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Metrel Eurocheck CS 2099 in the language / languages: English as an attachment in your email.

The manual is 1,33 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info