REVITIVE Ultrasound is intended for use in an electromagnetic environment in which
radiated RF disturbances are controlled. The user of REVITIVE Ultrasound can help prevent
electromagnetic interference by maintaining a minimum distance between portable and
mobile RF communications equipment (transmitters) and REVITIVE Ultrasound
as recommended below, according to the maximum output power of the
communications equipment.
Rated
maximum
output power
of transmitter
(W)
Separation distance according to frequency of transmitter (m)
150 KHz to 80
MHz
d= 3.5
V1
P
80 MHz to 800
MHz
d= 3.5
E1
P
800 MHz to 2.5 GHz
d= 7
E1
P
0.01 0.12 0.12 0.23
0.1 0.38 0.38 0.73
11.2 1.2 2.3
10 3.8 3.8 7.3
100 12 12 23
For transmitters rated at a maximum output power not listed above, the recommended
separation distance D in metres (M) can be estimated using the equation applicable to
the frequency of the transmitter, where P is the maximum output power rating of the
transmitter in watts (W) according to the transmitter manufacturer.
Note 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range
applies.
Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is
affected by absorption and reflection from structures, objects and people.
RECOMMENDED SEPARATION DISTANCES
BETWEEN PORTABLE AND MOBILE RF
COMMUNICATIONS EQUIPMENT AND EQUIPMENT
OR SYSTEMS THAT ARE NOT LIFE SUPPORTING.
Technical SpecificationsTechnical Specifications
REVITIVE Ultrasound is intended for use in the electromagnetic
environment specified below. The user of REVITIVE Ultrasound should assure that it is
used in such an environment.
Emission test Compliance Electromagnetic environment –
guidance.
RF emissions CISPR 11 Group 1
REVITIVE Ultrasound uses RF
energy only for its internal function.
Therefore, its RF emissions are very
low and are not likely to cause any
interference in nearby electronic
equipment.
RF emissions CISPR 11 Class B REVITIVE Ultrasound is suitable for
use in all establishments, including
domestic establishments and those
directly connected to the public
low-voltage power supply network
that supplies buildings used for
domestic purposes.
Harmonic
emissions IEC 61000-3-3 Class A
Voltage fluctuations/flicker
emissions IEC 61000-3-3 Complies
Power Supply:
Input: AC 100-240V, 50/60 Hz, 0.35A Output: DC 15V, 1.2A
Frequency: 1MHz (±10%)
Mode: 3 levels (Low, Medium, High)
Time: 5, 10 or 15 minutes with auto turn off
Power Output: (0.7W/cm2 (Low), 1.0W/cm2 (Medium), 1.2W/cm2 (High)
Duty Factor: 30% (Low), 40% (Medium), 50% (High)
RBN : < 8.0
Rated Output Power: 9.6W (±20%)
Effective Intensity: 2.4W/cm2 (±20%)
AER : 4 cm2
Beam Type: Collimated
Modulation Wave Shape: Pulsed
Operating Conditions:
Temperature: 5-40°C Humidity: 30-75%RH Air Pressure: 700-1060hPa
GUIDANCE AND MANUFACTURER’S DECLARATION -
ELECTROMAGNETIC EMISSIONS FOR
ALL EQUIPMENT AND SYSTEMS
3269_IFU02_14269060.indd 20-21 24/07/2017 11:59
11


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Revitive Ultrasound Therapy at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Revitive Ultrasound Therapy in the language / languages: English as an attachment in your email.

The manual is 5.67 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info