788491
11
Zoom out
Zoom in
Previous page
1/13
Next page
20 21
NOTE: Not intended to be sterilised. Not suitable for use in an OXYGEN
RICH ENVIRONMENT - if you are on oxygen we recommend you use your
REVITIVE Aerosure in a well-ventilated room.
Materials have been selected and tested to ensure no irritation or allergic
reaction in accordance with biocompatibility standards.
Technical Specifications
Product Model 3481AJ
Software version
(shown by V#.#) See underside of handle.
Charging Voltage 110v-240v
Power supply rating 30mA – 50mA
Battery & Battery Life Ni-MH / 500 cycles
Fuse T500mAH250V
Operating Environment Temperature 0°C to 40°C
Humidity 25 to 85% RH
Atmospheric pressure 86kPa - 106kPa
Storage and
Transportation Environment Temperature -18°C to 50°C
Humidity 25% to 85%RH
Protection rating against
ingress of water or
particulate matter.
IP45 (protected against small tools and
wires >1mm and low pressure water jets)
Protection against electric
shock
Internally powered ME equipment. Class
II (for charging unit).Type BF applied part
(valve head, handle and mouthpiece).
Mode of operation Continuous
Symbols
Contraindications. This describes situations
where you should not use the REVITIVE Aerosure
Warnings and cautions. Make sure you
understand these before using the REVITIVE
Aerosure
Type BF medical electrical equipment
Parts A, B and C (ref picture 1) are applied parts.
Medical electrical equipment double insulated
Complies with European Medical Devices
Directive (93/42/EEC)
The Waste Electrical and Electronic
Equipment Directive (WEEE Directive).
At the end of the product lifecycle, do not throw
this product into normal household waste, but
take it to a collection point for the recycling of
electronic equipment.
Legal manufacturer of the device
Consult instructions for use
Indicates the LOT number
Humidity and temperature limit
for storage and transportation
Indicates presence of NiMH battery.
(See page 20).
Indicates the item model reference number
LOT
NiMH
REF
-18°C
50°C
25%
85%
Medical electrical equipment needs special precautions regarding EMC for installation and
use according to the EMC information provided in this user manual.
Wireless communication equipment such as wireless home network devices, mobile phones,
cordless telephones and their base stations or walkie-talkies can affect this equipment and
should be kept at least a distance of 3.3m away from the equipment.
(Note: As indicated in Table 6 of IEC 60601-1-2:2007 for ME EQUIPMENT, a typical cell phone
with a maximum output power of 2 W yields d = 3,3 m at an IMMUNITY LEVEL of 3 V/m).
11


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Revitive Aerosure at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Revitive Aerosure in the language / languages: English as an attachment in your email.

The manual is 1.34 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info