Is the temperatures measured in the right ear different from that
measured in the left ear?
Among healthy people, there should be no significant difference in the
measurement results. Differences may be caused by the following reasons:
1) The infrared sensor is not inserted in the same way.
2) Measurement is not conducted by inserting the unit in a stable manner in
the same angle. Try to measure in either ear that can show the highest
measurement value constantly. In accordance with the “10 seconds mode,”
the unit detects the highest temperature. (Refer to paragraph 3.2).
The temperature shown is rather low.
1 The probe cover is dirty.
2 The infrared sensor is dirty.
3 You removed the unit from the ear before the measurement is finished.
4 The ear is cold. The temperature tends to indicate low when you use
an ice bag or an ice pack, or immediately after coming in from the
cold in winter.
5 The thermometer is not inserted deep enough in the ear.
6 The infrared sensor is not directed towards the eardrum.
Use the 10 seconds measurement mode (Refer to paragraph 3.2).
6 SPECIFICATIONS
Name: GentleTemp 510
®
Model: MC-510-E2
Voltage: 3V DC (lithium battery CR2032)
Power consumption: 0.05 W
Battery life: With a new battery approx. 5000 measurements
Sensor: Thermopile
Temperature indication: In 3 digits of +°C, every 0.1°F or °C
Precision: According to NEN-EN 12470- 5 Clinical
(Celsius) thermometers – Part 5 Performance of infra-red
ear thermometers (with maximum device)
34°C – 42.2°C: ±0.2°C
Based on the measurement of the standard black
body at the room temperature of 25°C (RH 50 %)
Precision: According to NEN-EN 12470- 5 Clinical
(Fahrenheit) thermometers – Part 5 Performance of infra-red
ear thermometers (with maximum device)
93.2°F – 108°F: ±0.3°C
Based on the measurement of the standard black
body at the room temperature of 77°F (RH 50 %)
Measurement range: 34°C - 42.2°C (93.2 - 108°F)
Weight: Approx. 50 g (1.8 oz) (including the battery)
External dimensions: Approx. 4.6 cm (B) X 9.3 cm (H) X 5.7 cm (D)
Operating environment: Ambient temperature; 10°C to 40°C (50°F to 104°F)
Relative humidity: 30 - 85%
Storage environment: Ambient temperature; -20°C-60°C (-4°F- 140°F)
Relative humidity: 10 - 95%
Accessories: Storage case, Instruction Manual, probe covers
(10 pieces)
A probe cover is already attached to the main unit.
Options: Probe cover exclusively for GentleTemp 510
®
series, MC-PROBE TYPE J
GentleTemp 510 English
13413 MC-510-E2_MC-510-E2 11/12/13 20:47 ページ 13
13


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Omron GentleTemp 510 at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Omron GentleTemp 510 in the language / languages: English as an attachment in your email.

The manual is 1,75 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

Others manual(s) of Omron GentleTemp 510

Omron GentleTemp 510 User Manual - German - 13 pages

Omron GentleTemp 510 User Manual - Dutch - 13 pages


The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info