Equipment Response Time Testing Summary
SpO
2
accuracy and low perfusion testing were conducted by Nonin Medical, Inc., as described below:
SpO
2
Accuracy Testing
SpO
2
accuracy testing is conducted during induced hypoxia studies on healthy, non-smoking, light- to dark-
skinned subjects during motion and no-motion conditions in an independent research laboratory. The measured
arterial hemoglobin saturation value (SpO
2
) of the sensors is compared to arterial hemoglobin oxygen (SaO
2
)
value, determined from blood samples with a laboratory co-oximeter. The accuracy of the sensors in comparison
to the co-oximeter samples measured over the SpO
2
range of 70 – 100%. Accuracy data is calculated using the
root-mean-squared (Arms value) for all subjects, per ISO 9919:2005, Medical Electrical Equipment–Particular
requirements for the basic safety and essential performance of pulse oximeter equipment for medical use.
Pulse Rate Motion Testing
This test measures pulse rate oximeter accuracy with motion artifact simulation introduced by a pulse oximeter
tester. This test determines whether the oximeter meets the criteria of ISO 9919:2005 for pulse rate during
simulated movement, tremor, and spike motions.
Low Perfusion Testing
This test uses a SpO
2
Simulator to provide a simulated pulse rate, with adjustable amplitude settings at various
SpO
2
levels for the oximeter to read. The oximeter must maintain accuracy in accordance with ISO 9919:2005
for heart rate and SpO
2
at the lowest obtainable pulse amplitude (0.3% modulation).
100
95
90
85
80
75
70
65
60
0.0
8.0
16.0
24.0
32.0
40.0
48.0
56.0
64.0
72.0
80.0
SaO
2
Reference
4 Beat Average
Time in seconds
SpO
2
Specific to this example: The response of the 4-beat average is 1.5 seconds.
7077-000-01 GO2 Achieve User Guide.indd 23-24 4/3/2009 10:14:33 AM
14


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Nonin Go2 at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Nonin Go2 in the language / languages: English as an attachment in your email.

The manual is 3,14 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

Others manual(s) of Nonin Go2

Nonin Go2 User Manual - German, Dutch, French, Italian - 2 pages


The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info