MI 3360 OmegaPAT XA Appendix E
152
No notifications mode
Instrument skips pre-test warnings (see chapter 4.5 Symbols and messages for more
information).
Parameters
State
On enables No notifications mode
Off disables No notifications mode
Appliance info
Instrument enables to automatically select the appliance type and add the appliance ID to the
Auto Sequence®.
Parameters
Repeat Setting
Repeat:
The same Appliance ID will be offered each time if the
same Auto Sequence® is carried out successively in a
loop.
Increment:
A four digit number will be added to the Appliance ID
and incremented each time if the same Auto
Sequence® is carried out successively in a loop.
Appliance type
Selects the type of the appliance (Appliance, Appliance_FD, Medical
Equip., Medical Equip. FD, Welding Equip., Welding Equip. FD)
Default Appliance ID
Enter default Appliance ID
Note
This flow command is active only if Auto Sequence® is started from the Auto
Sequence® Main menu.
Flow Protocol
This flow command controls commands for communication with external device for the control
of the flow of Auto Sequences®.
Parameters
Communication
settings
Selects the port for communication with the external device.
- RS232(PC)
- USB
Flow settings
Commands for communication with external device
(industrial PC for example)
Proceed
Send string:
Proceed
Description: If Proceed Flow setting is enabled,
implementation of Auto Sequences® on instrument can be
automatically controlled by external device by using
Proceed string command. Command works in parallel
with RUN key. Proceed Flow setting is disabled by default.
152


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Metrel MI 3360 - OmegaPAT XA at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Metrel MI 3360 - OmegaPAT XA in the language / languages: English as an attachment in your email.

The manual is 5.48 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

Others manual(s) of Metrel MI 3360 - OmegaPAT XA

Metrel MI 3360 - OmegaPAT XA User Manual - German - 172 pages

Metrel MI 3360 - OmegaPAT XA User Manual - Dutch - 79 pages


The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info