6) Mask for adults
7) Mask for children
8) Air filters
CLEANING AND DISINFECTING
Wash your hands well before beginning to clean and disinfect the accessories.
Do not wash the apparatus under the tap or by soaking. If it needs cleaning, just use a
cloth dampened with detergent (not abrasive and not solvent).
Device with enclosure not protected against the penetration of liquids.
To disinfect the accessories, it is possible to use denatured spirit or a hypochlorite solution
that can be easily purchased at any chemist’s shop.
Before and after each use, clean all nebuliser components (exception made for the air
tube) removing any medical product residuals or any dirt deposits.
Clean all components with lukewarm water and/or neutral soap; rinse them carefully
removing excess water and let them dry in the open air in a clean place. Never boil the
accessories.
To dry the humidity or condensate present in the air tube, switch on the device for 2/3
minutes letting only the tube connected; otherwise, let it dry at open air in a clean place.
Clean the external part of the tube with a wet cloth.
Replacement of the spare filter: Air filter has to be replaced every 500 hours of
operation or in any case when the same is rather dirty. It is suggested to replace the filer
every time that the ampoule is replaced (about every 6 months).
Do not wash the used filters to reuse them again.
To replace the filter, lift the filter holding cover, remove the filter to be replaced and insert
the spare one.
Then, insert the filter holding cover in its proper seat letting it adhere well to the surface.
Use only original Laica filters.
Cleaning of the ampoule: see ampoule instruction section
MICROBIC CONTAMINATION
In the presence of conditions carrying the risk of microbial infection and contamination,
personal use of the accessories and nebulizer only is advised (always consult a doctor).
PROBLEMS, CAUSES AND REMEDIES
The apparatus does not turn on:
ensure the plug is well into the socket.
ensure the switch is in the on position (I).
The apparatus is alight but does not nebulize:
ensure the air connection tube is not squashed or bent.
ensure the air suction filter is not blocked or dirty. If necessary, replace with a new
one.
ensure the required drug is inside the bulb.
make sure that the ampoule is not clogged; then, clean and disinfect the ampoule as
indicated in the manual. If the washing was not successful, replace the ampoule.
make sure that the nozzle is not badly mounted; press firmly with a finger on the
nozzle (cylindrical sleeve), which is positioned within the polycarbonate ampoule
bottom.
The appliance atomises slowly:
the medicinal product may be very oily. Dilute the medicinal product using
physiological saline.
The appliance is noisy after prolonged use:
Refer to the dealer.
The appliance does not work:
the appliance has worked close to heat sources or in environments with a temperature
exceeding 40°C. The thermal protection has tripped. Let the appliance cool for at
least 60 minutes and then switch on the appliance again.
N.B. if the apparatus still does not work correctly, contact the retailer.
HOW TO RETURN ITEMS
In compliance with the new European Regulations, LAICA lists here below
some essential points to preserve the hygiene of appliances and operators
using the same.
LAICA relies on the observance of these procedures to assure the hygiene
and health of all people operating in order to obtain quality and wellness.
Each appliance that shall be returned to LAICA shall be subject to hygienic controls before
technical service. In case of goods return for replacement, it is COMPULSORY to:
disinfect carefully the external part of the main unit using a cloth moistened with
denatured spirit or a hypochlorite solution.
disinfect the accessories plunging them in the same disinfecting solutions.
insert everything in a bag specifying “appliance with disinfected accessories”.
always specify the detected defect before returning the product.
TECHNICAL SPECIFICATION
Piston electrocompressor with thermal cut-out
13
Instructions and guarantee
GB
English
6


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Laica NE2007 at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Laica NE2007 in the language / languages: English, German, French, Spanish as an attachment in your email.

The manual is 1,74 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info