783854
13
Zoom out
Zoom in
Previous page
1/14
Next page
13 /14
ISO81060-2 2013,(Non-Invasive Sphygmomanometers - Part 2: Clinical Validation Of
Automated Measurement Type).
ELECTROMAGNETIC COMPATIBILITY INFORMATION
This product is applicable to the equipment and system requirements for the purpose of
receiving radio frequency energy for the purpose of the work, Bluetooth receive bandwidth
2M. This product can also be used to include RF transmitter equipment and system
requirements and emission frequency of 2.4GHz ISM band, Bluetooth modulation
types:GFSK, effective max radiated power: <4dBm
Table 1 - Emission
Phenomenon Compliance Electromagnetic environment
RF emissions CISPR 11
Group 1, Class B
Home healthcare environment
Harmonic distortion IEC 61000-3-2
Class A
Home healthcare environment
Vo l t a g e f l u c t u a t i o n s
and flicker
IEC 61000-3-3
Compliance
Home healthcare environment
Table 2 - Enclosure Port
Phenomenon Basic EMC standard Immunitytest levels
Home Healthcare Environment
Electrostatic
Discharge
IEC 61000-4-2 ±8 kV contact
±2kV, ±4kV, ±8kV, ±15kV air
Radiated RF EM field IEC 61000-4-3 10V/m
80MHz-2.7GHz
80% AM at 1kHz
Proximity fields from
RF wireless
communications
equipment
IEC 61000-4-3 Refer to table 3
Rated power
frequency magnetic
fields
IEC 61000-4-8 30A/m
50Hz or 60Hz
Table 3 Proximity fields from RF wireless communications
equipment
Test frequency
(MHz)
Band
(MHz)
Immunity test levels
Professional healthcare facility environment
385 380-390 Pulse modulation 18Hz, 27V/m
450 430-470 FM, ±5kHz deviation, 1kHz sine, 28V/m
710 704-787 Pulse modulation 217Hz, 9V/m
745
780
13


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for iHealth BP5S - Neo at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of iHealth BP5S - Neo in the language / languages: English as an attachment in your email.

The manual is 0.87 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

Others manual(s) of iHealth BP5S - Neo

iHealth BP5S - Neo User Manual - German - 26 pages

iHealth BP5S - Neo User Manual - Dutch - 25 pages

iHealth BP5S - Neo User Manual - French - 26 pages

iHealth BP5S - Neo User Manual - Spanish - 26 pages


The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info