5.1.2 Recharging the Batteries
Attention!
!
Use only the charger (Z507A) which is offered as an op-
tional accessory for this instrument to charge the batter-
ies inserted in the instrument.
Make sure that the following conditions have been fulfilled be-
fore connecting the charger to the charging socket:
– rechargeable batteries have been inserted with correct
polarity, no normal batteries
– The test instrument has been disconnected from the
measuring circuit at all poles
– The instrument must remain off during charging.
Recharging of the batteries begins as soon as the charger is con-
nected to the mains and to the charging socket (see figure 2.1).
Charging takes approximately 5 hours and 30 minutes (integrated
safety timer) if the batteries have been fully depleted.
Safety Precautions
Do not attempt to recharge alkaline batteries: they may leak,
explode etc. The test instrument may be severely damaged or
destroyed as a result.
After initially charging new batteries and after rechargeable
batteries have not been used for a lengthy period of time (sev-
eral months), operating hours after charging may be signifi-
cantly less than usual. If this is the case, repeat the charging
procedure several times.
Autonomous, intelligent charging stations execute charging/
discharging cycles of this sort automatically (see instructions
included with the charging station). This procedure increases
the capacity of the batteries, thus making longer periods of
operation possible between charging cycles.
If no improvement is achieved in this way, one or more of the
rechargeable batteries may no longer fulfill the original specifi-
cations. If this is the case, the defective rechargeable battery
should be identified, e.g. with the help of the voltage measure-
ment, and replaced.
Battery capacity is gradually reduced as a result of long and
frequent use. When you notice that this is the case, all of the
rechargeable batteries should be replaced.
5.2 Cleaning
Use a soft cloth and soapy water for cleaning. Do not place the
test instrument back into service until its surface is completely dry.
Attention!
!
Do not use cleaning agents which contain benzine or al-
cohol! Prevent liquids from penetrating into the test in-
strument’s interior.
5.3 Recalibration
The measuring tasks performed with your instrument, and the
stressing it’s subjected to, influence aging of its components and
may result in deviation from the specified levels of accuracy.
In the case of strict measuring accuracy requirements, as well as
in the event of use at construction sites with frequent stress due
to transport and considerable temperature fluctuation, we recom-
mend a relatively short calibration interval of once per year. If your
instrument is used primarily in the laboratory and indoors without
considerable climatic or mechanical stressing, a calibration inter-
val of once every 2 to 3 years is sufficient as a rule.
During recalibration at an accredited calibration laboratory
(DIN EN ISO/IEC 17025), deviations from traceable standards
demonstrated by your measuring instrument are documented.
Ascertained deviations are used to correct display values during
later use of the instrument.
We would be happy to perform DAkkS or factory calibration for
you at our calibration laboratory. Further information is available at
our website:
www.gossenmetrawatt.com ( Company DAkkS Calibration
Center or FAQs General – Calibration Questions and
Answers).
Recalibration of your instrument at regular intervals is essential for
the fulfillment of requirements according to quality management
systems per DIN EN ISO 9001.
* Examination of the specification, as well as adjustment, are not included
in calibration. However, in the case of our own products, any required
adjustment is performed and adherence to the specification is con-
firmed.
6 Repair and Replacement Parts Service,
Calibration Center and Rental
Instrument Service
If required please contact:
GMC-I Service GmbH
Service Center
Thomas-Mann-Str. 20
90471 Nürnberg, Germany
Phone: +49 911 817718-0
Fax: +49 911 817718-253
e-mail service@gossenmetrawatt.com
www.gmci-service.com
This address is only valid in Germany. Please contact our repre-
sentatives or subsidiaries for service in other countries.
7 Product Support
If required please contact:
GMC-I Messtechnik GmbH
Product Support Hotline
Phone: +49-911-8602-0
Fax: +49 911 8602-709
e-mail support@gossenmetrawatt.com
8


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Gossen Metrawatt METRALINE ISO-CHECK at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Gossen Metrawatt METRALINE ISO-CHECK in the language / languages: English as an attachment in your email.

The manual is 0,39 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

Others manual(s) of Gossen Metrawatt METRALINE ISO-CHECK

Gossen Metrawatt METRALINE ISO-CHECK User Manual - Dutch - 8 pages


The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info