783888
8
Zoom out
Zoom in
Previous page
1/8
Next page
32
Manufacturers declaration-electromagnetic immunity
Test specications for ENCLOSURE PORT IMMUNITY to RF wireless communications equip-
ment
The device is intended for use in the electromagnetic environment (for home healthcare) specied
below. The customer or the user of the device should assure that it is used in such an environment
Test frequency
(MHz)
Band a)
(MHz)
Service a) Modulation b) Maximum
power (W)
Distance
(m)
IMMUNITY
TEST LEVEL
(V/m)
Compliance
LEVEL (V/m)
(for home
healthcare)
385 380 – 390 TETRA 400 Pulse
modulation b)
18 Hz
1,8 0,3 27 27
450 430 – 470 GMRS 460,
FRS 460
FM c) ±5 kHz
deviation 1 kHz
sine
2 0,3 28 28
710 704 – 787 LTE Band
13, 17
Pulse
modulation b)
217 Hz
0,2 0,3 9 9
745
780
810 800 – 960 GSM 800/900,
TETRA 800,
iDEN 820,
CDMA 850,
LTE Band 5
Pulse
modulation b)
18 Hz
2 0,3 28 28
870
930
1 720 1700 –
1990
GSM 1800;
CDMA 1900;
GSM 1900;
DECT;
LTE Band 1, 3,
4, 25; UMTS
Pulse
modulation b)
217 Hz
2 0,3 28 28
1 845
1 970
2 450 2400 –
2570
Bluetooth,
WLAN,
802.11 b/g/n,
RFID 2450,
LTE Band 7
Pulse
modulation b)
217 Hz
2 0,3 28 28
5 240 5100 –
5800
WLAN
802.11 a/n
Pulse
modulation b)
217 Hz
0,2 0,3 9 9
5 500
5 785
NOTE If necessary to achieve the IMMUNITY TEST LEVEL, the distance between the transmitting antenna and the ME EQUIPMENT
or ME SYSTEM may be reduced to 1 m. The 1 m test distance is permitted by IEC 61000-4-3.
a) For some services, only the uplink frequencies are included.
b) The carrier shall be modulated using a 50 % duty cycle square wave signal.
c) As an alternative to FM modulation, 50 % pulse modulation at 18 Hz may be used because while it does not represent actual
modulation, it would be worst case.
8


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Gima TH889 at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Gima TH889 in the language / languages: German as an attachment in your email.

The manual is 0.5 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

Others manual(s) of Gima TH889

Gima TH889 User Manual - English - 8 pages

Gima TH889 User Manual - French - 8 pages


The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info