121
Immunity
test IEC 60601
test level Compliance
Level Electromagnetic environment
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
3 Vrms
da 150 kHz
80 MHz
3 V/m
da 80 MHz
2,5
GHz
N/A
3 V/m
Portable and mobile RF communications
equipment should be used no closer to
any part of this device, including cables,
than the recommended separation dis-
tance calculated from the equation appli-
cable to the frequency of the transmitter..
Recommended separation distance
d= 1.2 √P
d= 1.2 √P
80MHz to 800MHz
d= 2.3 √P
800MHz to 2.5GHz
Where P is the maximum output power
rating of the transmitter in watts (W) ac-
cording to the transmitter manufacturer
and d is the recommended separation
distance in meters (m).
Field strengths from xed RF transmitters,
as determined by an electromagnetic site
survey,
a
should be less than the compli-
ance level in each frequency range.
b
Interference may occur in the
vicinity of equipment
marked with
the following symbol:
a: Field strengths from xed transmitters, such as base stations for radio (cellular /
cordless) telephones and land mobile radios, amateur radio, AM and FM radio broad-
cast and TV broadcast cannot be predicted theoretically with accuracy.
To assess the electromagnetic environment due to xed RF transmitters, and electro-
magnetic site survey should be considered. If the measured eld strength in the loca-
tion in which the device is used exceeds the applicable RF compliance level above,
the device should be observed to verify normal operation. If abnormal performance is
observed, additional measures may be necessary, such as reorienting or relocating
this device.
b: Over the frequency range 150 kHz to 80 MHz, eld strengths should be less than
3V/m.
18


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Gima OXY 4 at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Gima OXY 4 in the language / languages: German as an attachment in your email.

The manual is 0.47 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

Others manual(s) of Gima OXY 4

Gima OXY 4 User Manual - English - 19 pages

Gima OXY 4 User Manual - French - 19 pages


The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info