Puls oximeter
Introduction:
The Fysic FPO-11 is a pulse Oximeter which
measures the percentage of the amount of oxy-
gen found in the blood.
Intended use:
The Pulse Oximeter is a non-invasive device
intended for spot checking of functional oxygen
saturation of arterial hemoglobin (SpO2) and
pulse rate (PR). This portable device is indicated
for use in adult patients in clinical institution and
home environments.
Measuring principle:
By shining an infra-red light through the skin, the
reected light will be compared with the emitted
light which will in turn analyze the color of the red
blood cells. This colour measured will reect the
amount of oxygen in the blood.
Do not use:
Do not use the FPO-11 with an irregular heart
beat (Atrial Fibrillation).
Installation:
The FPO-11 comes with two included batteries, a
wrist strap and a protective case.
Install batteries:
1. Slide the battery cover o󰀨 of the FPO-11
2. Insert the two
supplied batte-
ries, be sure to
take note of the positive and negative poles
(+ and -)
3. Place the battery cover back onto the FPO-11
Fastening the wristband:
1. Push the thin cord through the
hole on the FPO-11 and then
make a loop from the cord.
2. Put the thick part of the strap through the loop.
3. Pull the thick part of the cord all the way
through the loop tugging slightly to tighten it.
Use:
Preparations & tips:
Make sure that you are familiar with the ope-
ration of the pulse Oximeter.
Avoid bright lights.
Remove articial nails and/or nail polish.
Wash your hands before using.
Make sure your hands are warm to ensure a
good blood circulation.
Sit quietly before and during the taking of the
measurement.
Measurement:
1. Open the clip and place your nger on the
FPO-11 with the nail facing upwards, as
shown on the adjacent picture; use your mid-
dle nger preferably pushing it gently towards
the back of the Oximeter.
2. Turn on the FPO-11 by pressing the
on/o󰀨 button.
3. If necessary, the reading direction
of the display can be changed by
pressing the on/o󰀨 button repea-
tedly.
4. Do not move your nger during the measurement.
5. Readings in the display:
- The “% SpO2” shows the percentage of
oxygen in the blood.
- The “PRbpm” gives the pulse rate of the
pulse per minute.
- The PI (Perfusion Index) is the ratio of the
pulsatile blood ow to the non-pulsatile
blood ow. The PI’s values range from
0.02% for very weak pulses to 20% for
strong pulses.
- Depending on the selected reading di-
rection, a pulsating bar graph shows the
strength of the heartbeat or the heart rate
can be seen at the bottom of the display.
6. The FPO-11 will turn o󰀨 15 seconds after the
clip has been taken o󰀨 the nger.
Setting an alarm:
You can set alarm tones to give alert signals when
the level of the oxygen in your blood, or the pulse
rate exceeds the adjustable upper or lower values.
1. Turn on the FPO-11 by quickly pressing the
on/o󰀨 button.
2. Hold down the on/o󰀨 button for more than one
second to open the settings menu.
3. Now release the on/o󰀨 button. You will see
one of the following statements:
The line with the star at the end is the active line.
Press quickly on the on/o󰀨 button to move the
star to the next line.
Press and hold the on/o󰀨 button to change
the setting of the current line.
Limit Setup *
SPO2 Hi 100
SPO2 Lo 94
PR Hi 130
PR Lo 50
+/- +
Exit
Remind Setup *
Sound Reminder on
Beep o󰀨
Demo o󰀨
Restore Ok
Brightness 4
Exit
GB
1


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Fysic FPO-11 at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Fysic FPO-11 in the language / languages: English as an attachment in your email.

The manual is 0.37 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

Others manual(s) of Fysic FPO-11

Fysic FPO-11 User Manual - Dutch - 2 pages

Fysic FPO-11 User Manual - French - 2 pages


The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info