Calibrations and Custom Test Standards
NVP Calibration
6
6-3
NVP Calibration
The Determine Cable NVP function allows you to determine the NVP (nominal
velocity of propagation) for a known length of cable and save the value for
additional measurements of unknown lengths of the same type of cable. The
calibrated NVP value applies only to the selected test standard. You can also enter
a known NVP value or use the Set to Default softkey to set the NVP to
the default value, which is the NVP for a typical sample of the selected cable type.
To determine a cable’s NVP value, proceed as follows:
Notes
To determine a cable’s NVP, you should use a cable at least 15m
(50 ft) long. The recommended length is 30m (100 ft).
NVP is determined for the cable pair with the shortest electrical
delay.
1. Turn the rotary switch to SPECIAL FUNCTIONS.
2. Use D to highlight Determine Cable NVP.
3. Press E.
4. As prompted by the display, connect a known length of the appropriate cable
to the test tool. On Model DSP-2000, connect the cable to the CABLE TEST
jack.
5. Verify that no cables are connected to the unused cable connector or DB9
connector on the test tool.
6. Press T.
7. Use D or U to change the displayed length to the known length of the
attached cable. As you adjust the length measurement, the NVP value changes
automatically. The range for NVP values is 50.0% to 99.9%.
8. When the displayed length matches the known length of the cable, press S
to store the new NVP value and return to the main Special Functions menu.
To exit the screen without saving the new NVP value, press e.
If the message CABLE FAULT DETECTED appears, verify that the selected
cable type is correct, check the cable for proper pin connections, or calibrate with
a different cable.
109

Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Fluke DSP-2000 at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Fluke DSP-2000 in the language / languages: English as an attachment in your email.

The manual is 1,66 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info