8808A
Users Manual
A-8
Making Low-Level Current Measurements
There are many applications where obtaining the utmost accuracy in low-level current
measurements is critical. For example, determining the leakage current of a battery
operated device in its standby mode is critical in determining the time before battery re-
charge is needed. Traditional multimeters make these measurements using a shunt
method shown in Figure A-3. The shunt resistor converts the current to be measured to a
voltage, which is called the burden voltage. Since the internal impedance of the current
source is in parallel with the shunt resistor, the current flowing in the shunt resistor is less
than the actual value, thus causing an error.
A
V
OUT
I
IN
R
SHUNT
R
Internal
eue28.eps
Figure A-3. Shunt Method of Low-Level Current Measurement
Another method of low-current measurement is the feedback resistor method shown in
Figure A-4. The feedback resistor converts the current to be measured into a voltage. The
high gain operational amplifier forces the burden voltage to approximately zero and thus
reduces the error associated with the simple shunt measurement approach. The zero
burden voltage measurement method used in the Meter gives a more accurate
measurement of low-level (leakage) currents.
VOUT
IIN
RFeedback
RInternal
A
eue29.eps
Figure A-4. Zero Burden Voltage Low-Level Current Measurement
98


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Fluke 8808A at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Fluke 8808A in the language / languages: English as an attachment in your email.

The manual is 1,96 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

Others manual(s) of Fluke 8808A

Fluke 8808A User Manual - German - 102 pages


The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info