17
Test Certificates
Seite 1
/
2
EG-Baumusterprüfbescheinigungen ohne Unterschrift und ohne Stempel haben keine Gültigkeit.
Diese EG-Baumusterprüfbescheinigung darf nur unverändert weiterverbreitet werden.
Auszüge oder Änderungen bedürfen der Genehmigung der Prüf- und Zertifizierungsstelle ZELM Ex
Prüf- und Zertifizierungsstelle ZELM Ex Siekgraben 56 D-38124 Braunschweig
-perten
für
-pl osionsschutz
Prüf- und Zertifizier ungsstelle
ZE LM
(1)
EC type approval test certificate
(2) Devices and protection systems for proper designated use
in areas with an increased risk of explosion – Directive 94/9/EC
(3) EC type approval test certificate number
ZELM 02 ATEX 0120 X
(4) Device: Explosion-protected mA calibrator 707Ex
(5) Manufacturer: ecom instruments GmbH
(6) Address: D-97959 Assamstadt
(7) The design of this device and its various approved embodiments are defined in the attachment to
this type approval test certificate.
(8) As the nominated body no. 0820 in accordance with article 9 of the EC council directive dated
March 23, 1994 (94/9/EC), the Testing and Certification Body ZELM Ex certifies conformance with
the basic health and safety requirements for the design and construction of devices and protection
systems for proper designated use in areas with an increased risk of explosion in accordance with
Appendix II of the directive.
The results of the test are documented in the confidential test report no. ZELM Ex 0510217162.
(9) The basic health and safety requirements are met by virtue of conformance with
EN 50 014: 1997+A1+A2 EN 50 020: 1994
(10) If the certification number is followed by an “X”, then this indicates that special conditions exist for
the safe operation of the device. These special conditions are contained in the attachment to this
certificate.
(11) This EC type approval test certificate only refers to the construction/design, checking and testing of
the specified device or protection system in accordance with directive 94/9/EC. Further require-
ments contained in this directive may apply with regard to the manufacturing process and the supply
of the device or protection system. Such requirements are not covered by this certification.
(12) The device must be labelled with the following information:
II 2 G EEx ia IIC T4
Certification body: ZELM Brunswick, 28.02.2003
Dipl.-Ing. Harald Zelm
Seite 2
/
2
EG-Baumusterprüfbescheinigungen ohne Unterschrift und ohne Stempel haben keine Gültigkeit.
Diese EG-Baumusterprüfbescheinigung darf nur unverändert weiterverbreitet werden.
Auszüge oder Änderungen bedürfen der Genehmigung der Prüf- und Zertifizierungsstelle ZELM Ex
Prüf- und Zertifizierungsstelle ZELM Ex Siekgraben 56 D-38124 Braunschweig
-perten
für
-plosionsschutz
Prüf- und Zertifizier ungsstelle
ZE LM
(13)
Attachment
(14)
EC type approval test certificate ZELM 02ATEX 0120 X
(15) Description of the device
The explosion-protected mA calibrator 707Ex serves as a compact source and measuring instru-
ment for mobile measurements of current loops (0...24 mA) and DC voltages (up to 28 V) in areas
with an increased risk of explosion as well as in areas with no increased risk of explosion.
The device is equipped with a power supply in the form of a battery of type 6LR61. This battery
must only be changed outside the area with increased risk of explosion.
Within the ex-area the device must be carried in the designated holster.
Electrical data
Power supply circuit from an integral battery, EEx ia IIC
The internal circuits are intrinsically safe.
Measuring circuit type of ignition protection: intrinsically safe, EEx ia IIC
for connection to certified intrinsically safe circuits
Maximum values:
a) Current measuring circuit, Uo = 27.6 V DC
active (0...24 mA) (linear characteristic) Io = 96.9 mA
Max. perm. external capacitance Co = 76 nF
Max. perm. external inductance Lo = 2.5 mH
b) Voltage measurement (0...28 V)Ui=30V DC
Ii = 24 mA
Effective internal capacitance Ci = 10 nF
The effective internal inductance is negligibly small.
Permitted ambient temperature -10 °C to +50 °C
Measurements on non-intrinsically safe circuits
Measurement inputs Maximum values: U = 30 V DC
I=24mA
Note:
The operating instructions must be followed.
The battery must be changed outside the ex-area.
Within the ex-area the device must be carried in the designated holster.
Attachment to EC type approval test certificate ZELM 02 ATEX 0120 X
(16) Test report no.
ZELM Ex 0510217162
(17) Special conditions
The battery must be changed outside the ex-area. Only the battery type(s) listed in the operating
instructions must be used.
Within the ex-area the device must be carried in the designated holster.
(18) Basic health and safety requirements
satisfied by virtue of the standards
Certification body:
ZELM Brunswick, 28.02.2003
Dipl.-Ing. Harald Zelm
21


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Fluke 707Ex at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Fluke 707Ex in the language / languages: English as an attachment in your email.

The manual is 0,75 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

Others manual(s) of Fluke 707Ex

Fluke 707Ex User Manual - German - 22 pages

Fluke 707Ex User Manual - Dutch - 22 pages


The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info