True-rms Digital Multimeters
Changing Meter Setup Options
47
The primary display indicates the input signals pulse width in
milliseconds. The frequency of the signal is displayed in the
secondary display. The mini-measurement display indicates the
volts or amps value of the input signal. The bar graph tracks the
volts or amps value of the signal and not the pulse width value.
The pulse width polarity is displayed to the right of the duty cycle
value. J indicates a positive pulse width and K indicates a
negative pulse. To change the polarity, press the softkey labeled
J K. The polarity indicator changes to the opposite polarity.
Changing Meter Setup Options
The Meter has a number of preset features such as date and
time formats, backlight and battery save mode timeouts, and the
displayed language. These variables are referred to as Meter
setup options. Many setup options affect general Meter
operations and are active in all functions. Others are limited to
one function or group of functions.
Access to the setup options is always available through the
softkey labeled Setup. Information about the Meter, such as
serial number, model, for example. is also accessed through the
setup menu.
Resetting Meter Setup Options
The Meter’s setup options can be reset to default values through
the setup menu. Open the setup menu by pressing the softkey
labeled Setup. Position the menu selector next to the menu item
labeled Reset and press the softkey labeled Setup. A message
will appear asking to confirm the reset action. Press the softkey
labeled OK to perform the reset.
Note
A setup reset also resets the temperature offset and
dBm reference to their default value.
In addition to resetting the setup variables, pressing the softkey
labeled Meter will also clear all saved measurement screens,
MIN MAX screens, peak screens, and recording records. The
Meter’s clock is reset to a default value as well.
Setting Display Contrast
Meter display contrast can be adjusted through the Meter’s setup
menu. Open the setup menu by pressing the softkey labeled
Setup and position the menu selector next to the menu item
labeled Contrast. Pressing the softkey labeled + (F1) increases
display contrast, while the softkey labeled (F2) decreases
contrast.
Contrast can also be set through the 7 and 8 buttons when not
being used to move between menu selections.
Setting the Meter’s Language
The Meter comes from the factory with the display language set
to English. To select another language, open the setup menu by
pressing the softkey labeled Setup. Move the menu selector next
to the menu item labeled Display. Next, press the softkey
labeled Format (F2) to open the format menu. If not already
selected, move the menu selector to the left of the menu item
labeled Language and press the softkey labeled Edit. The
currently selected language becomes highlighted and f appears
to the right of the language. Use 7 and 8 to scroll through the
available languages, then press the softkey labeled OK to set the
Meter’s display language. Press the softkey labeled Close to
return to normal Meter operation.
57


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Fluke 289 at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Fluke 289 in the language / languages: English as an attachment in your email.

The manual is 2,15 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

Others manual(s) of Fluke 289

Fluke 289 User Manual - German - 92 pages

Fluke 289 User Manual - Dutch - 90 pages


The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info