763957
52
Zoom out
Zoom in
Previous page
1/88
Next page
287/289
Users Manual
42
W Caution
Placing the probes across (in parallel with) a
powered circuit when a lead is plugged into a
current terminal can damage the circuit you are
testing and blow the Meter's fuse. This can happen
because the resistance through the Meter's current
terminals is very low, so the Meter acts like a short
circuit.
The following are tips for current measurements:
A current meter drops a small voltage across itself, which might
affect circuit operation. You can calculate this burden voltage
using the values listed in the specifications under Burden Voltage
(A, mA, μA).
The Meter’s current functions have a number of modes which
provide more details about a current signal. Pressing the softkey
labeled Menu opens a menu of items that can be used to modify
the basic current measurement. Refer to the appropriate section
in this manual to learn more about each menu item.
To clear all modes and return to the basic ac or dc current
measurement, press the softkey labeled Menu. Move the menu
selector to the item labeled AC,DC. Press the softkey labeled AC
to clear all functions and modes and make basic ac current
measurements, or DC for basic dc current measurements.
Measuring Frequency
Frequency is the number of cycles a signal completes each
second. The meter measures the frequency of a voltage or
current signal by counting the number of times the signal crosses
a threshold level within a specified period of time.
Figure 23 highlights the functions that allow frequency
measurements.
est21.eps
Figure 23. Functions Allowing Frequency
Measurement
The Meter autoranges to one of five frequency ranges:
99.999 Hz, 999.99 Hz, 9.9999 kHz, 99.999 kHz, and 999.99 kHz.
Figure 24 shows a typical frequency display. Pressing R
controls the input range of the primary function (volts or amps)
and not the frequency range.
To measure frequency, rotate the switch to one of the primary
functions allowing frequency measurements highlighted in
Figure 23. Press the softkey labeled Menu and move the menu
52


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Fluke 289 at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Fluke 289 in the language / languages: English as an attachment in your email.

The manual is 2,15 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

Others manual(s) of Fluke 289

Fluke 289 User Manual - German - 92 pages

Fluke 289 User Manual - Dutch - 90 pages


The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info