Table 6. Accuracy Specifications
Function Range Resolution Accuracy
± ([% of Reading] + [Counts]) Model
DC Millivolts 600.0 mV 0.1 mV 0.5 % + 2 110, 114, 115, 117
DC Volts
6.000 V 0.001 V
0.5 % + 2 110, 114, 115, 11760.00 V 0.01 V
600.0 V 0.1 V
DC, 45 to 500 Hz 500 Hz to 1 kHz
Auto-V LoZ[1] True-rms 600.0 V 0.1 V 2.0 % + 3 4.0 % + 3 114, 117
kCHEK[4]
6.000 V 0.001 V
2.0 % + 3 4.0 % + 3 11360.00 V 0.01 V
600.0 V 0.1 V
45 to 500 Hz 500 Hz to 1 kHz
AC millivolts[1] True-rms 600.0 mV 0.1 mV 1.0 % + 3 2.0 % + 3 110, 114, 115, 117
AC Volts[1] True-rms
6.000 V 0.001 V
1.0 % + 3 2.0 % + 3 110, 114, 115, 11760.00 V 0.01 V
600.0 V 0.1 V
Continuity[5]
600 Ω1 ΩBeeper on <20 Ω, off >250 Ω. Detects
opens or shorts of 500 μs or longer.
110, 114, 115, 117
--- --- 113
Ohms[5]
600.0 Ω0.1 Ω0.9 % + 2 110, 113, 114, 115,
117
6.000 kΩ0.001 kΩ0.9 % + 1
60.00 kΩ0.01 kΩ0.9 % + 1
600.0 kΩ0.1 kΩ0.9 % + 1
110, 114, 115, 117
6.000 0Ω0.001 MΩ0.9 % + 1
40.00 MΩ0.01 MΩ5.0 % + 2
Diode Test[5] 2.000 V 0.001 V 0.9 % + 2 115, 117
2.0 % + 3 113
Capacitance[5]
1000 nF 1 nF 1.9 % + 2
113, 115, 117
10.00 μF0.01 μF1.9 % + 2
100.0 μF0.1 μF1.9 % + 2
9999 μF1 μF100 μF - 1000 μF: 1.9 % +2
>1000 μF: 5 % + 20
Lo-Z Capacitance
(Power-up option) 1 nF to 500 μF 10 % + 2 typical 115, 117
AC Amps True-rms[1]
(45 Hz to 500 Hz)
6.000 A 0.001 A 1.5 % + 3 115, 117
10.00 A[3] 0.01 A
DC Amps 6.000 A 0.001 A 1.0 % + 3 115, 117
10.00 A[3] 0.01 A
20


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Fluke 113 at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Fluke 113 in the language / languages: English as an attachment in your email.

The manual is 1,21 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

Others manual(s) of Fluke 113

Fluke 113 User Manual - German - 22 pages

Fluke 113 User Manual - Dutch - 22 pages


The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info