Compex Fit - User Manual
Immunity
test
Test level IEC
60601
Observance
level
Electromagnetic
environment - guide
Electrostatic
discharge
(DES)
CEI 61000-4-2
± 6 kV at the
contact
± 8 kV in air
± 6 kV at the
contact
± 8 kV in air
Floors must be wood, concrete
or ceramic tiles. If floors
are covered with synthetic
material, the relative humidity
must be maintained at a
minimum of 30%.
Rapid electrical
transients in
bursts of noise
CEI 61000-4-4
± 2 kV for power
supply lines
± 1 kV for input/
output lines
Not applicable
System battery-
powered
The quality of the power
supply should be that of a
typical commercial or hospital
environment.
Surge currents
CEI 61000-4-5
± 1 kV differential
mode
± 2 kV common
mode
Not applicable
System battery-
powered
The quality of the power
supply should be that of a
typical commercial or hospital
environment.
Voltage
trough, short
power cuts
and voltage
variations on
power
supply input
lines
CEI 61000-4-11
< 5% U
T
(trough
>95% of U
T
) for 0,5
cycle
40% U
T
(trough 60%
of U
T
) for 5 cycles
70% U
T
(trough 30%
of U
T
) for 25 cycles
<5% U
T
(trough
>95% of U
T
) for 5 sec
Not applicable
System battery-
powered
The quality of the power
supply should be that of a
typical commercial or hospital
environment. If the Compex
Fit user requires continuous
operation during mains power
cuts, we recommend that the
Compex Fit is powered by a
UPS or a battery.
Magnetic field
at mains
frequency
(50/60 Hz)
CEI 61000-4-8
3 A/m Magnetic fields at the mains
frequency should be at the
level of a representative site
located in a typical commercial
or hospital environment.
Compex Fit is designed for use in the electromagnetic environment
stipulated below. The buyer or user of the Compex Fit must ensure it is used
in this recommended environment.
NOTE: U
T
is the AC supply voltage before application of the test level.
RECOMMENDATIONS AND DECLARATION BY THE
MANUFACTURER - ELECTROMAGNETIC IMMUNITY
171


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Compex Fit at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Compex Fit in the language / languages: English, German, French, Italian, Spanish as an attachment in your email.

The manual is 4,17 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info