TROUBLESHOOTING
1918
WARNING: No modification of this equipment is allowed.
PROBLEM SYMPTOM CHECK THIS REMEDY
No power
Low
batteries
Error
massage
Display is dim or
will not light up.
Batteries are exhausted.
Replace with new batteries
Insert the batteries
correctly
Replace with new batteries
Batteries are inserted
incorrectly.
Show on
the display
Batteries are low.
E 1 shows
Data Communication
Failure.
Please check below items:
1.Bluetooth
®
Smart is ON.
2.Both devices are within
the transmission distance of
Bluetooth
®
Smart
E 2 shows The cuff is very tight
Refasten the cuff and then
measure again.
E 3 shows
The pressure of the
cuff is excess.
Relax for a moment
and then measure again.
The monitor detected
motion while measuring.
movement can affect the
measurement.Relax for a
moment and then
measure again.
Measure incorrectly.
Relax for a moment and
then measure again.
A calibration error
occurred.
Retake the measurement.
If the problem persists,
contact the retailer or our
customer service
department for further
assistance.Refer to the
warranty for contact
information and return
instructions.
E 10 or
E 11 shows
E 20 or
E 21 shows
Eexx,shows on
the display.
Data Transmission
Failure.
Bluetooth
®
Smart is OFF.
The device is too far
away from your mobile.
Press and hold START·
STOP button to start
manual data transmission.
Turn ON Bluetooth
®
Smart
and try again.
Keep the distance of Blood
Pressure Monitor and the
mobile within 2 to 3 meters.
SPECIFICATIONS
Power supply
Battery Powered Mode:
6V (4 x AAA-size alkaline-battery)
AC Adaptor Powered Mode:
100-240 V~, 50-60 Hz, 400 Ma
(Can be supplied by AC adaptor model
UE08WCP-060100SPA only!)
Display mode
Blue LCD with White Backlight
V.A. = 141mm(L) x 36mm(W)
Measurement mode
Oscillographic testing mode
Measurement range
Pressure: 0-40kpa(0~300mmHg)
pulse value:(40-199)times/minute
Accuracy
Pressure:
15 -25 within ±0.4 kPa (3 mm Hg)
10 -40 (out of 15 -25 ) within ±0.7 kPa
(5 mm Hg); Pulse Value: ±5%
Working condition
Temperature:10
-40
Relative Humidity 15%-90%RH
Atmospheric Pressure: 80-105 kPa
Storage & transportation
condition
Measurement perimeter
of the upper arm
About 22cm-32cm
Net Weight
Approx.340g(Excluding the dry cells)
External dimensions
Attachment
Approx.180*100*40mm
4*AAA alkaline batteries,user manual
Mode of operation
Continuous operation
Degree of protection
Type BF applied part
IP Classification
IP22
Temperature:-20-60
Relative Humidity 10%-93%RH
Atmospheric Pressure: 50-106 kPa
Software Version
V01
Device Classification
Battery Powered Mode: Internally Powered ME Equipment
AC Adaptor Powered Mode: Class II ME Equipment
11


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for BodyGauge TMB-1018-BT at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of BodyGauge TMB-1018-BT in the language / languages: English as an attachment in your email.

The manual is 1,07 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info