Beurer GL44 mmol/L 33
Storage/
transport temperature
Temperature: +2 °C – +30 °C
Relative humidity: < 90 %
Operating ranges Temperature: +10 °C – +40 °C
Relative humidity: < 90 % non-condensing
Measuring range, glucose Glucose: 1.1 – 35.0 mmol/L
Blood sample capillary whole blood
Amount of blood 0.6 microlitres
Blood glucose measuring duration Approx. 5 seconds
Calibration Plasma
Test procedure Amperometric bio sensor
Using the device Suitable for self-testing
System function test Upon every switch-on
EMC
This device complies with the European standard EN 61326 and is subject to particular precautions
with regard to electromagnetic compatibility. Please note that portable and mobile HF communica-
tion systems may interfere with this unit. For more details, please contact our customer service at
the address indicated.
Test strip functionality
Test strips enable a quantitative measurement of the glucose level in fresh whole blood. When the gap
for taking blood comes into contact with a drop of blood, it is automatically filled by simple capillary
action. The blood is sucked into the absorbing gap on the test strip and the blood glucose monitor
measures the blood glucose level in the blood.
The test is based on the measurement of an electric current that is generated by the chemical reaction
of the glucose with the enzyme glucose dehydrogenase (Aspergillus sp.) on the strip.
During the reaction, a mediator transports electrons through the electrode surface and so gener-
ates a current.
The blood glucose monitor analyses this current. The current flow is proportional to the glucose con-
tent in the blood sample. The results are shown on the blood glucose monitor display. Only a small
amount of blood is required (0.6 microlitres) and measurement duration is approx. five seconds. The
test strip detects blood glucose values from 1.1 to 35.0 mmol/L.
Chemical components of the test strip sensor
FAD glucose dehydrogenase 6 %
Potassium ferricyanide 56 %
Non-reactive components 38 %
Control solution functionality
The control solution contains a fixed amount of glucose that reacts with the test strip. A test with con-
trol solution is similar to a blood test. However, control solution is used instead of blood. The measured
result using control solution must be within the result range. This range is printed on each test strip box.
33


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Beurer GL 44 mmol L at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Beurer GL 44 mmol L in the language / languages: English as an attachment in your email.

The manual is 6,14 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

Others manual(s) of Beurer GL 44 mmol L

Beurer GL 44 mmol L User Manual - German - 40 pages

Beurer GL 44 mmol L User Manual - Dutch - 40 pages

Beurer GL 44 mmol L User Manual - French - 40 pages


The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info