Table 2 ‐ IMMUNITY TEST LEVELS
Phenomenon Basic EMC standard IMMUNITY TEST LEVELS
Electrostatic discharge IEC 61000‐4‐2 ±8 kV contact
±2 kV, ±4 kV, ±8 kV, ±15 kV air for en-
closure port
Radiated RF electroma-
gnetic fields
IEC 61000‐4‐3 3 V/m
80 MHz to 2.7 GHz
80 % AM at 1 kHz for enclosure port
Proximity fields from RF
wireless communications
equipment
IEC 61000‐4‐3 See table 3
Rated power frequency
magnetic fields
IEC 61000‐4‐8 30 A/m
50 Hz and 60Hz for enclosure port
Table 3 - Test specifications for ENCLOSURE PORT IMMUNITY to RF wireless communi-
cations device
Test
frequency
(MHz)
Band
(MHz)
Service Modulation Maximum
power (W)
Distance
(m)
IMMUNITY
TEST LEVEL
(V/m)
385 380 to 390 TETRA 400 Pulse
modulation
18 Hz
1.8 0.3 27
450 430 to 470 GMRS 460,
FRS 460
FM
±5 kHz
deviation
1kHz sine
2 0.3 28
710 704 to 787 LTE Band
13, 17
Pulse
modulation
217 Hz
0.2 0.3 9
745
780
230


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Beurer BM 93 at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Beurer BM 93 in the language / languages: English, German, French, Italian, Polish, Turkish, Spanish as an attachment in your email.

The manual is 12,2 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info