Short Instructions
BENNING PV 1-1
Important information
m
Before using the BENNING PV 1-1 please read the detailed operat-
ing manual (CD-ROM) carefully.
The BENNING PV 1-1 should only be used by suitably trained per-
sonnel.
c
The connection to the PV generator is made exclusively in accord-
ance with the connection figure of the operating manual.
Disconnect not required safety tests leads from the BENNING PV 1-1.
c
The PV module/PV string must not exceed the maximum open-
circuit voltage of 1000 V DC, the maximum short-circuit current of
15 A and the maximum DC power (P = U x I) of 10 kW.
According to DIN EN 62446, the measurements have to be made for
each PV string.
Measurements made at PV strings which are connected in parallel
might involve an overload of the BENNING PV 1-1.
c
Disconnect the BENNING PV 1-1 from the test sample directly after
the test.
c
Do not touch the measuring probes! During insulating resistance
measurements, high electric currents might be applied to the meas-
uring probes.
c
Do not touch any metal parts of the test object during measure-
ment.
c
The PV generator must be isolated from the electric power supply!
Neither the positive nor the negative pole of the PV generator must
be earthed!
c
Via the 4 mm safety test leads, voltage measurements on mains
supply circuits are possible. Via the 4 mm test sockets, the
BENNING PV 1-1 must be used only in electric circuits of overvolt-
age category III with max. 300 V AC/DC for phase-to-earth measure-
ments. For this please disconnect the PV 1-1 measuring leads from
the PV test sockets before measuring.
c
Before starting the unit, always check it for signs of damage. Do not
use a damaged BENNING PV 1-1!
m
Only use safety measuring leads, which are supplied with the
BENNING PV 1-1.
c
The BENNING PV 1-1 is intended for making measurements under
dry ambient conditions only.
Switching the device ON/OFF
Press the
R
PE
-key and the
Auto
-key simultaneously to switch the device ON or OFF.
Without pressing a key, the device switches OFF automatically after approx. 1
minute (APO, Auto Power-Off).
Measured value memory (200 display screens)
-Store
Store all measurements currently on the LC display. In the RE-
CALL mode, the measuring results are called in reverse order.
-Recall
Recall the stored measured values on the LC display.
+
Clear all results from memory.
Reading out the measured value memory via the USB interface
1. Install the driver and download program from the CD-ROM once.
2. Disconnect all measuring leads from the BENNING PV 1-1.
3.
Connect the BENNING PV 1-1 to your PC by means of the USB connecting cable.
4. Start the download program, select the COM port and click “Download”.
5.
Switch on the BENNING PV 1-1 and press the -key for approx. 2 seconds.
6. Measured value download will start.
Preset limiting values
V
ISO
Limiting value of insulating resistance
250 V 0.5
500 V/1000 V 1.0
Measuring range
Function Range
R
PE
0 Ω - 199 Ω/30 V - 440 V
Vo/c 5 V - 1000 V
Is/c 0.5 A - 15 A
R
ISO
0.2 MΩ - 199 MΩ
I
AC/DC
0.1 A - 40 A
Setting the date and time
1. Turn off the BENNING PV 1-1.
2. Press and hold the
-
key and then press simultaneously the
R
PE
-key and
the
Auto
-key of BENNING PV 1-1.
3. The date format and time format is shown as follows:
MM.DD = month (1-12). Day (1-31)
YYYY = year
HH.mm = hours (0-23).minutes (0-59)
SS = seconds (0-59)
4. Press the
R
PE
-key to select a date field and time field
5. A blinking field shows that this field can be set.
6. With the -key and the
-
key, the value increases or decreases. With
each change, the second field is set to zero.
7. Turn off the device to save the setting.
Setting the Auto Power Off (APO, Auto-Power Off)
1. Turn off the BENNING PV 1-1.
2. Press and hold the
Ω
NULL
-key and then press simultaneously the
R
PE
-key and
the
Auto
-key of BENNING PV 1-1. Keep the
Ω
NULL
-key pressed.
3. The LC display shows the first line “OFF” and in the second row the shutdown
in minutes.
4. Each press of the
V
ISO
-key increases the shutdown time by one minute to 10
minutes.
5. Then release the
Ω
NULL
-key to save the setting.
Radio connection to BENNING SUN 2 - wireless SUN link
The BENNING PV 1-1 is able to receive the measured values (insolation, PV mo-
dule/ambient temperature and date/time stamp) of the insolation and temperature
measuring instrument BENNING SUN 2 (optional) via radio connection. For this
purpose, the BENNING PV 1-1 has to be coupled with the BENNING SUN 2 once.
Typical radio range of the BENNING SUN 2 in open space: approx. 30 m
Buildings/metal structures or interfering signals can reduce the radio range.
Coupling with insolation/temperature measuring instrument BENNING SUN 2
1. Remove all electronic devices in direct vicinity
2. Switch the BENNING PV 1-1 and the BENNING SUN 2 off.
3. Press and hold the two ON/OFF keys of the BENNING SUN 2.
4. Press and simultaneously hold the
R
PE
-key and the
Auto
-key of the
BENNING PV 1-1.
5. The BENNING PV 1-1 indicates the successful coupling by means of an
acoustic signal and by displaying the serial no. of the BENNING SUN 2
6. The "W/m
2
" symbol is shown on the LC display of the BENNING PV 1-1.
Decoupling from insolation/temperature measuring instrument BENNING SUN 2
1. Remove all electronic devices in direct vicinity.
2. Switch the BENNING PV 1-1 off.
3. Press and hold the
R
PE
-key and the
Auto
-key of the BENNING PV 1-1 for approx.
10 seconds.
4. The BENNING PV 1-1 indicates the decoupling from the BENNING SUN 2 by
means of an acoustic signal and by clearing the LC display.
5. The "R
PE
/Ω" symbol is shown on the LC display of the BENNING PV 1-1.
Activating/deactivating the radio transmission of the BENNING SUN 2
1. Couple the BENNING PV 1-1 with the BENNING SUN 2.
2. To activate/deactivate the radio transmission, press and hold the -key of
the BENNING SUN 2 and simultaneously press the -key. A flashing triangle
above the -key shows that the radio transmission has been activated.
3. If the BENNING PV 1-1 is within the radio range of the BENNING SUN 2,
the measured insolation value (W/m
2
) is shown on the LC display of the
BENNING PV 1-1.
4.
Besides the electric variables (Vo/c, Is/c, R
ISO
), AUTO measurement by means
of the BENNING PV 1-1 additionally measures the insolation, the module and
ambient temperature as well as the date/time stamp of the BENNING SUN 2.
5. If the BENNING PV 1-1 is outside the radio range of the BENNING SUN 2, the
"W/m
2
" on the LC display starts flashing. Moreover, "_ _ _ _" is shown on the
LC display, if the measured insolation value is outside the measuring range.
Note:
If the BENNING PV 1-1 does not receive any radio signal from the BENNING SUN 2,
the display indications are stored with the date/time stamp of the BENNING PV 1-1.
Error codes
Error codes Remedy
The internal fuse has blown. Refer to chapter 9.5 in the opera-
ting instructions for details.
The electronic components of the BENNING PV 1-1 have
reached the maximum admissible temperature. Disconnect the
BENNING PV 1-1 from the object to be measured and let it
cool down.
The DC short-circuit current has the maximum value of 15 A.
The measurement has been stopped.
The DC open circuit voltage has exceeded the maximum value
of 1000 V. The measurement has been stopped.
Other error codes see detailed user guide on CD-ROM.
09/ 2012
BENNING PV 1-1
109/ 2012
BENNING PV 1-1
609/ 2012
BENNING PV 1-1
5
3


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Benning PV 1.1 at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Benning PV 1.1 in the language / languages: Dutch, French, Italian, Spanish as an attachment in your email.

The manual is 19,25 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

Others manual(s) of Benning PV 1.1

Benning PV 1.1 User Manual - English, German - 29 pages


The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info